KENCO ENGINEERING COMPANY
P.O. BOX 470426 TULSA, OK 74147-0426 PHONE: (918) 663-4406 FAX: (918) 663-4480
www.kenco-eng.com e-mail: [email protected]
SMARTSONIC INTEGRAL ACOUSTIC WAVE LEVEL / OPEN CHANNEL FLOW TRANSMITTER
INSTALLATION / OPERATION INSTRUCTIONS
TABLE OF CONTENTS
1. GENERAL DESCRIPTION ............................................................. 2
2. PRINCIPLE OF OPERATION ......................................................... 2
3. MODEL DESCRIPTION .................................................................. 2
4. INSTALLATION .............................................................................. 2
5. SPECIFICATIONS .......................................................................... 2
6. DIMENSIONS.................................................................................. 3
7. MOUNTING GUIDELINES .............................................................. 3
8. MOUNTING INSTRUCTIONS ......................................................... 4
9. WIRING ........................................................................................... 4
Wiring Connections....................................................................... 5
Loop Resistance Table (Intrinsically Safe).................................. 5
Loop Resistance Table (Non-Intrinsically Safe).......................... 5
10. CALIBRATION................................................................................ 6
Display / Keypad............................................................................ 6
11. INITIAL SET-UP PROGRAM .......................................................... 7
Initial Set-up Program Menu......................................................... 7
Five Initial Programs (Pr.20 thru Pr.24) ....................................... 8
12. STANDARD PROGRAMS .............................................................. 11
Standard Program Menu............................................................... 11
Programs (Pr.01 thru Pr.10).......................................................... 12
13. OPEN CHANNEL FLOW MEASUREMENT ................................... 21
Program Pr.00 (Flow Measurement) ............................................ 21
Flume / Weir Styles and Dimensions ........................................... 22
Flumes / Weirs – U.S. Standard.................................................... 23
Flumes / Weirs – European........................................................... 26
14. TROUBLESHOOTING .................................................................... 29
15. APPENDIX A (Gas Factor Table) .................................................. 30
Vista de pagina 0
1 2 3 4 5 6 ... 29 30

Indice de contenidos

Pagina 1 - KENCO ENGINEERING COMPANY

Page 1 KENCO ENGINEERING COMPANY P.O. BOX 470426 TULSA, OK 74147-0426 ● PHONE: (918) 663-4406 FAX: (918) 663-4480 www.kenco-eng.com e-mail: info@ken

Pagina 2

Page 10 Initial Set-up Program Pr.21 (Flow Measurement Units) This program is only available if you selected Flow in Program 24. This is where you se

Pagina 3

Page 11 STANDARD PROGRAMS Standard Program Menu FlowMeasurementResetTankHeightInterferingSignals4mA20mAResponseSpeedWorkingAreaLevel / DistanceAtmosph

Pagina 4

Page 12 Reset Program The Pr.01 program enables you to do a reset, refreshing the SmartSonic measurement reading, while not changing data saved

Pagina 5

Page 13 Press / Action Display Description and simultaneously Enter the Program Menu One Time Cycle to Program #2 Enter P

Pagina 6 - Next button

Page 14 Function 00.16 (22mA Error Signal) This function allows you to define if “Near Zone” and “Lost Echo” are allowable errors, and if the result i

Pagina 7

Page 15 Each reading (scan distance) taken using the Pr.03 program is stored as an interfering signal, until a readin

Pagina 8

Page 16 Temporary display while SmartSonic searches for Interfering Signals After a brief pause For Example: Distance to Interfering Signal Sav

Pagina 9

Page 17 and For Example: Change the value to represent the desired 4mA Point in either: Level, Distance, or Flo

Pagina 10

Page 18 and For Example: Change the value to represent the desired 2mA Point in either: Level, Distance, or Flo

Pagina 11

Page 19 Working Area The Pr.07 program allows you to add a distance range that exceeds the tank’s height, thus enabling accurate readi

Pagina 12

Page 2 GENERAL DESCRIPTION The SmartSonic Acoustic Wave Transmitter is designed for high accuracy level measurement of a variety of liquids. The Smar

Pagina 13

Page 20 For Example: Displays the current distance/level mode setting and Used to cycle between Distance (d000) and Level (L000)

Pagina 14

Page 21 Restoring the Default Settings The Pr.10 program allows you to clear programs Pr.00, Pr.02 – Pr.09, and reverts to Factory Sett

Pagina 15

Page 22 Press / Action Display Description and simultaneously Enter the Program Menu One Time Cycle to Program #00 For E

Pagina 16

Page 23 Flumes / Weirs – U.S. Standard (dimensions are in inches unless otherwise indicated) Type 1 (X) – Rectangular Suppressed Sharp-Crested Weir

Pagina 17

Page 24 Type 4 (X) – V-notch (Triangular) Sharp-Crested Weir Type 5 (X) - Parshall Flume Type 6 (X) – Palmer-Bowlus Flume

Pagina 18

Page 25 Type 7 (X) – H-Flume Type 8 (X) – Leopold-Lagco Flume Code (FF) Flume Size Measurement Point 01 6 1.96 02 9

Pagina 19

Page 26 Flumes / Weirs – European (dimensions are in centimeters unless otherwise indicated) Type 1 (X) – Rectangular Suppressed Sharp-Crested Weir

Pagina 20

Page 27 Type 4 (X) – V-notch (Triangular) Sharp-Crested Weir Type 5 (X) – Khafagi-Venturi Flume Type 6 (X) – Pa

Pagina 21

Page 28 Type 7 (X) – Palmer-Bowlus Flume Type 8 (X) – H-Flume Type 9 (X) – Neyrpic-Venturi Flume / Long Base Weir

Pagina 22

Page 29 TROUBLESHOOTING This section describes how to resolve problems that may occur when calibrating / operating the SmartSonic. If the following i

Pagina 23

Page 3 DIMENSIONS MOUNTING GUIDELINES 1.46 4.25 2.55 2.16 1.00 7.63 3.75 0.40 35º 1.18 The transmitter must be mounted in such

Pagina 24

Page 30 Appendix A. – Gas Factor Table Atmosphere (Gas) Factor Atmosphere (Gas) Factor Acetic Acid 0.62 Ethane 0.90 Acetone 0.63 Ethanol 0.71

Pagina 25

Page 4 MOUNTING INSTRUCTIONS Precautions: • Ensure that the transmitter is mounted in an area that meets the stated temperature, pressure and technic

Pagina 26

Page 5 Wiring Connections Before starting installation procedures in hazardous areas, insure that all power sources have been turned off and locked o

Pagina 27

Page 6 CALIBRATION This section explains how to set up and calibrate the SmartSonic for accurate measurement monitoring. SmartSonic is supplied with

Pagina 28

Page 7 NOTE: Within some programs, the digits in the display can be individually modified. This is indicated by a flashing digit. In this case the E

Pagina 29

Page 8 Five Initial Programs While the Pr.02 Program is normally used to set the Tank Height, there is a special code that is used to enter this Initi

Pagina 30

Page 9 Initial Set-up Program Pr.23 (Measurement Range) This program allows you to set the transmitter to either Standard or Short Range. The Standar